Wheatstone's Bridge
WHEATSTONE'S BRIDGE, an electrical instrument which consists of six conductors, joining four points, of such a character that when an electromotive force is applied in one branch the absence of a current in another branch (called the conjugate branch) establishes a relation between the resistance of the four others by which we can determine the value of the resistance in one of these, that of the others being assumed to be known. This arrangement was not invented by Sir Charles Wheatstone although it bears his name and is commonly attributed to him, and was employed by him in some of his electrical researches but by S. H. Christie, in I833. 1 The arrangement of the six conductors is diagrammatically represented in fig. i. In one of these branches is placed a battery B and in another a galvanometer G ; the four other resistances are denoted by the letters P, Q, R, S. The circuits in which the battery and galvanometer are placed are called conjugate circuits, ^and the circuits P, Q, R, and S are I called the arms of the bridge, the /branches P and Q being called the ratio arms and S the measuring arm. The circuit in which the galvanometer is placed is the bridge circuit. Keys are inserted in the battery and galvanometer circuits to open or close them at pleasure. The resistance forming the four arms of the bridge can be so adjusted that if these resistances have values denoted by P, Q, R, and S, then when P : Q : : R : S, the current in the galvanometer circuit will be zero when an electromotive force is applied in the battery circuit.
To prove this statement, let the conductors P, Q. R, S., be arranged in a lozenge shape, as in fig. I. Let E be the electromotive force in the battery circuit, and let (x+y) be the current through the resistance P, y the current through the resistance Q and 2 that through B. Then by G. R. Kirchhoff's laws (see ELECTROKINETICS) we have the current equations, (P+G+R) (x+y)-Gy-Rz=O (Q+G + S)} G(x + y)-Sz = (R+S+B) z-R(* + ?) -Sy = E Rearranging the terms and solving for x (the current through the galvanometer), we obtain * = (PS-RQ)E/A, where A is a complex expression, involving the resistances P, Q, R, S, G, and B, which does not concern us. Hence when * = o, P :Q = R :S and the value of R can be determined in terms of P, Q and S.
In the practical instrument the three arms of the bridge P, Q, and S are generally composed of coils of wire contained in a box, whilst R is the resistance the value of which is to be determined. This last resistance is connected to the other three with the addition 1 See Wheatstone's Scientific Papers, p. 129.
of a galvanometer and a battery connected up as shown in the diagram. The operation of determining the value of the resistance R therefore consists in altering the ratio of the three resistances P, Q, and S, until the galvanometer indicates no current through it when the battery circuit is completed or closed by the key. In one form of Wheatstone's Bridge, known as the scries patterrr plug-resistance bridge, or Post Office pattern, the two ratio arms, P and Q, each consist of a series of coils of wire, viz. two i-ohm coils, two lo-ohm coils, two loo-ohm coils and two looo-ohm coils, which are joined up FIG. 2. Standard Wheatstone's Bridge.
in series in the order, 1000, 100, 10, I ; I, 10, 100, 1000, the junction* between each pair being connected to brass blocks, a series of which are mounted upon an ebonite slab that forms the lid of the box. The blocks are bored out with a hole partly in one block and partly in the other (see fig. 2) so that they can be connected by accurately fitting conical plugs. When the blocks are interconnected by the plugs all the coils are short-circuited ; but if the plug or plugs are taken out, then a current flowing from one end of the series to the other is compelled to pass through the corresponding coils. In series with this set of coils is another set, S, which forms a measuring arm, the resistances of which are generally 1,2,3,4, IO > 2 9> 3> 4> IOO > 2OO > 300, 400, looo, 2000, 3000, 4000 ohms. The junction between each pair of coils is connected as above described to a block, the blocks being interconnected by plugs all of which are made interchangeable. Another form of Wheatstone's Bridge, shown in fig. 2, is known as the dial pattern. Ten brass blocks are arranged parallel to or around another brass block, and by means of a plug which fits into holes bored partly out of the common block and partly out of the surrounding blocks, any one of the latter can be connected with the common one. A series of nine equal resistances, say I-ohm coils, or nine loo-ohm coils, are joined in between these circumferential blocks (fig. 3). It will be seen that if a plug is placed so as to Connect FIG. 3.- -Diagram showing Connexions of a Dial and Plug pattern, Wheatstone's Bridge.
any outside block with the central block, the current can only pass from the zero outer block to the central block by passing through a certain number of the resistance coils. Hence according to the magnitude of each coil the total resistance may be made anything from i to Q, 10 to 90, or 100 to 900 ohms, etc. Three or four of the " dials " thus composed are arranged side by side, the brass blocks being mounted on a slab of ebonite and the coils contained in the box underneath, and they are so joined up that the central block of one dial is connected to the outside block of the next marked O. This arrangement forms the measuring arm of the bridge, the ratio arms being constructed on the series plug pattern just described. A bridge of this pattern has the advantage that the insertion or removal of a plug in the measuring arm does not tend to tighten or loosen all the TIM of the plugs; moreover, there are fewer plugs to manipulate, and each plug is occupied. The resistance coils themselves are generally wound on brass or copper bobbins, with silk-covered mangamn wire, which should first be aged by heating for about ten hours to a temperature of 140 C., to remove the slight tendency to change in resistivity which would otherwise present itself.
For the accurate comparison of resistance coils it is usual to make use of the Matthicssen and Hockin bridge, and to employ the method of differential comparison due to G. Carey Foster. 1 On a board is stretched a uniform metallic wire a b, generally of platinum silver. The ends of this wire are connected to copper blocks, which themi are connected to a series of four resistance coils, A, B, and P, Q (fig. 4). A and B are the coils to be compared, P and Q are two other coils of convenient value. Over the stretched wire moves a contact maker S, which makes contact with it at any desired point, the position of which can be ascertained by means of an underlying scale. A battery C of two or three cells is connected to the extremities of the slide wire, and the sensitive galvanometer G is connected in between the contact-maker and the junction between the coils P and Q. The observer begins by moving the slider until the fjulvanometer shows no current. The position of the coils A and B is then interchanged, and a fresh balance in position on the bridge is obtained. It is then easily shown that the difference between the resistance of the coils A and B is equal to the resistance of the length of the slide wire intercepted between the two places at which the balance was found in the two observations.
Let the balance be supposed to be attained, and let x be the position of the slider on the wire, so that x and lx are the two sections of the slide wire, then the relation between the resistance is Next, let the position of A and B be interchanged, and the slide-wire reading be x' ; then (B+*')/ (A +/-*') = P/Q.
Hence it follows that A B=x' x, or the difference of the resistances of the coils A and B is equal to the resistance of that length of the slide wire between the two points where balance is obtained.
Various plans have been suggested for effecting the rapid interchange of the two coils A and B; one of the most convenient was designed by J. A. Fleming in 1880, and has been since used by the British Association Committee on Electrical Units for making comparison between standard coils with great accuracy (see Phil. Mag., 1880, and Proc. Phys. Soc., 1879). In all very exact resistance measurements the chief difficulty, however, is not to determine the resistance of a coil, but to determine the temperature of the coil at the time when the resistance measurement is made. The difficulty is 1 " On a Modified Form of Wheatstone's Bridge, and Methods of Measuring Small Resistances," by Professor G. Carey Foster. Proc. Soc. Tel. Eng. (1872), i.
caused by the fact that the coil is heated by the current used to measure its resistance, which thus alters in value. In accurate comparisons, therefore, it is necessary that the coils to be compared should be immersed in melting ice, and that sufficient time should be allowed to elapse between the measurements for the heat generated in the coil to be removed.
The standard resistance coil employed as a means of comparison by which to regulate and check other coils consists of a wire, generally of mangamn or platinum silver, insulated with silk and wound on a brass cylinder (fig. 5). This is soldered to two thick terminal rods of copper, and the coil is enclosed in a water-tight brass cylinder so that it can be placed in water, or preferably in paraffin oil, and brought to any required temperature. In the form of standard coil recommended by the Berlin Rcichsanstalt the coil is immersed in an insulating oil which is kept stirred by means of a small electric motor during the time of making the measurement. The temperature of the oil can best be ascertained by means of a platinum resistance thermometer.
For the measurement of low resistances a modification of the Wlieatstone's bridge devised by Lord Kelvin is employed. The Kelvin bridge consists of nine conductors joining six points, and in one practical form is known as a Kelvin and Varley slide. Modifications of the ordinary Wheatstone's bridge for very accurate measurements have been devised by H. L. Callendar and by Callendar and E. H. Griffiths (see G. M. Clark, the Electrician, 38, p. 747). A useful bridge method for measurement of low resistances has been given by R. T. Housman (the Electrician, 40, p. 300, 1897). These and numerous modifications of the Wheatstone's bridge will be found described in I. A. Fleming's Handbook for the Electrical Laboratory and TestingRoom, vol. i. (1903).
REFERENCES: F.E. Smith, "OnMethodsof High Precision for the Comparison of Resistances," Appendix to the Report of the British Association Committee on Electrical Standards, British Association Report (York, 1906), or the Electrician, 57, p. 976 (1906); C. V. Drysdale, " Resistance Coils and Comparisons," British Association Report (Leicester, 1907), or the Electrician, 57, p. 955 (1907), and 60, p. 20 (1907); J. A. Fleming, " A Form of Resistance Balance for Comparing Standard Coils," Phil. Mag. (February, 1880); "A Design for a Standard of Electrical Resistance," Phil. Mag. (January 1889); G. Aspinall Parr, Electrical Measuring Instruments (1903); W. H. Price, The Practical Measurement of Resistance; A. Gray, Absolute Measurements in Electricity and Magnetism (1900); Rollo Appleyard, " The Conductometer," Proc. Phys. Soc. London, 19, p. 29 (1903); also Proc. Inst. Civ. Eng. 154 (1903); and Proc. Phys. Soc., London, 17, p. 685 (1901). (J. A. F.)